JULABO FP45-HE Refrigerated and Heating Circulators User Manual
Page 35

HE
35
The started program
A
B
C1
C2
D1
D2
After the start the program will indicate the currently calculated setpoint in line
2
S XX.XX. The value increases within the time period >TSLICE< until the
target temperature >SETPNT< of the section is reached.
If the time period in a section is set to „0“, the next section will not begin until
the target temperature has been reached.
Use the edit keys
to scroll to line 3. The display changes
approximately every 4 seconds between the current section (STEP XX) and
the
A remaining time of the section
B remaining time of the program
C current bath temperature
I xxx.xx - internal actual value or
E xxx.xx – external actual value
D RUN – the program has started or
PAUSE – the progress of the program has been interrupted by pressing
the
key. While the time is stopped the temperature will constantly
remain at the last calculated setpoint
Continue with the
key.
Termination / Interruption of a program
The program can be terminated any time by pressing the key
.
In case of power failure the program is interrupted.
The circulator switches to –OFF- .
If the AUTOSTART-function is activated the programmer starts again at
the point in time approx. 5 minutes prior to the interruption. However, an
uncontrolled change of the bath temperature has occurred.