In Data Exchange mode, the DP master class 1 cyclically transmits data from and
to all slaves that are connected to the bus.
In this document, data transmitted from the slave to the master are called "input
data" and data transmitted from the master to the slave are called "output data".
The input and output data of the BPG400-SP / HPG400-SP have two logic parts:
1) the parameter channel
2) the process data channel
There is a number of standard telegrams, consisting of:
a) the parameter channel only
b) the process data channel only
c) both, the parameter and process data channel
The parameter channels allows masters without Profibus DPV1 to access device
specific parameters that are not part of the normal cyclic data telegram. For mas-
ters with Profibus DPV1, no parameter channel is required.
The input data (transmitted by the BPG400-SP / HPG400-SP) consist of the
8 bytes of the parameter channel (if there is a parameter channel in the standard
telegram) and of 5 … 7 bytes of process data depending on the selected standard
telegram.
Byte
Byte
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
Parameter channel
Process data
PKE
IND res.
PWE
Where: PKE
=
Parameter Signature Value
Reading or writing com-
mand and definition of the slot
IND
=
Sub Index
Index No. of the index to be
read (
→ "Block Model")
res.
=
reserved
PWE
=
Process Value
Value to be read or written
Assignment of the block ele-
ments to the slot indices
2.2 Structure of the Cyclic
Data Telegrams in Data
Exchange Mode
Input data