tira36e1-a (0310) BPG/HPG400 v1.cp
Two data types are supported: Float and Integer16.
In cyclic data exchange, the data type cannot be modified. The data type setting
can only be modified when the gauge is in the IDLE status. By defining the
configuration data for cyclic data exchange (selection of standard telegrams), the
data type used in the selected standard telegram is taken over. All settings
previously made in acyclic data traffic are thus overwritten.
If the data type is set in one instance, that data type setting applies to all instances.
Likewise, when a standard telegram is selected, the data type used by that
standard telegram will be valid for all instances.
Coding
Data type
3
Integer16
8
Float
The gauge supports the following pressure units:
Coding
Data type
1001
COUNTS
1301
Torr
1302
mTorr (Micron)
1308
mbar
1309
Pascal
For safety reasons, it is not possible to change the pressure unit while
the gauge is cyclically interchanging data with a DP/V0 master.
The data unit setting can only be modified when the gauge is in the IDLE
status.
In cyclic data traffic, the data unit must be set in the User Parameter
Data. All settings previously made in acyclic data traffic are overwritten
(
→ section "User Parameter Data").
If the data unit is set in one instance, that data unit setting applies to all
instances. Likewise, the data unit setting made in the User Parameter
Data is valid for all instances.
This parameter indicates that the pressure reading is within a valid range, which
means that:
• The gauge is in the EXECUTING status.
• There is no error (ID 26, 27 or 28 of the Device Block)
• The measured value is lower than the overrange value and higher than the
underrange value.
If this value is set to zero, the pressure reading is not valid. In such a
case, either check Exception Status (ID 26, Device Block) to find out
whether there is an error or check One Of N Status Extension (ID 120,
One Of N Vacuum Gauge Transducer Block) to find out whether the
measured value is out of the specified measuring range (overrange or
underrange mode).
This parameter contains the valid maximum value of the pressure reading
(1000 mbar) in the currently selected data unit and data type.
Pressure unit
Full scale
COUNTS
31 000
Torr
750.06168
mTorr (Micron)
750 061.68
mbar
1 000
Pascal
100 000
3.2.2.3 Data Type
3.2.2.4 Data Unit
3.2.2.5 Reading Valid
3.2.2.6 Full Scale
BGP400-SP