35
ID
Name
Structure Data type
Bytes Access
Store Initial Value
14 Hot Cathode
Block State
Command
Rec
1
1_W
2_W
FALSE
101 Block Type
Simple
Octet
string
4
2_R
16
102 Status
Extension
Simple
UINT8
1
1_R
2_R
V
0
103 Sensor
Alarm
Struct
Array of
2 bytes
2
1_R
2_R
V
0
104 Sensor
Warning
Struct
Array of
2 bytes
2
1_R
2_R
V
0
105 Emission
Status
Simple
Boolean
1
1_R/W
2_R/W
V
0
106 Emission
Current
Simple
FLOAT
4
1_R/W
2_R/W
NV
BPG400-SP
only
109 Degas
Status
Simple
Boolean
1
1_R
2_R
V
BPG400-SP
only
According to the table in Appendix a, the Block Type ID has the value "16".
This parameter indicates that the overrange or underrange of the Hot Cathode Ion
Gauge instance is exceeded.
Bit7
Bit6
Bit5
Bit4
Bit3 Bit2
Bit1
Bit0
0
0
0
0
0
Underrange
Exceeded
Overrange
Exceeded
Reading
Invalid
If the instance is operated in its overrange or underrange, the corresponding bit
and additionally the bit "Reading Invalid" is set.
If an error occurs, the bit "Reading Invalid" as well as the corresponding error bits
in Device Block (ID 26, 27, 28 Device Block) are set.
This parameter indicates the detectable errors occurring in connection with the
BA/HP measuring part. The present implementation allows detection of one error.
Bit7
Bit6
Bit5
Bit4
Bit3
Bit2 Bit1
Bit0
Byte 0
0
0
0
0
0
0
0
0
Byte 1
0
0
0
0
0
0
Electronics Failure
1)
0
1)
Electronics Failure includes a sensor error.
The Sensor Alarm bits defined here are copied into the Device Block ID 27 in the
"Device Exception Detail Alarm" range of the BA measuring part.
This parameter indicates the detectable warnings occurring in connection with the
BA/HP measuring part. The present implementation allows detection of one
warning.
Bit7
Bit6
Bit5
Bit4
Bit3
Bit2
Bit1
Bit0
Byte 0
0
0
0
0
0
0
0
0
Byte 1
0
0
0
0
0
0
Electronics Warning
0
The Sensor Warning bits defined here are copied into the Device Block ID 28 in the
"Device Exception Detail Warning" range of the BA/HP measuring part.
3.3.3 Hot Cathode Ion Gauge
Transducer Block /
SLOT 1
3.3.3.1 Block Type
3.3.3.2 Status Extension
3.3.3.3 Sensor Alarm
3.3.3.4 Sensor Warning