Maintenance Plan

*) every 2 years regardless of operation hours

Required maintenance

Operation hours/Years

Repair

level

Assembly

UL5000

1500

4000

8000

16000

Part no.

1/4

1

2

Vacuum system

Scroll pump IWATA

Exchange the scrollmodule

X

III

200000021R

ISP 500

Scroll pump Varian

Exchange the Tip Seal

X

1

III

200001671

TS 620

Exchange the scrollmodule

X

III

200001665R

TMH 071

Replace the lubricant reservoir

X*

II u. III

200000577

Exchange of bearings

X

III

Valve bloc

Clean the valves, replace seals for valves

2

X

III

200000594

Take apart the valve bloc and clean it

2

X

III

200000593

Replace filters for vent- and purge line

1

X

1

I, II ,III

200000683

Adjust the Pirani

X

III

Silencer

Exchange the silencer

X

1

I, II, III

20099183

Electric

Fans assembly

Clean fans at chassis plate and side wall
e.g by pressurized air

1

X

1

I

Exchange spare filter cell for fans chassis
plate

1

X

1

I

20000685