24

Installation

iina

74

e 0

2

.f

m

te

chn

ica

l ha

nd

boo

k

(1

40

8)

2.3.2.1

Accessories

The following accessories may be connected to this port

Fig. 2-6

/1:

external venting valve

sniffer probe

Contact pins 1 and 3 are fused with a 0.8 A slow-blow fuse. The amount of power
which can be drawn is limited to 10 W. The contacts are numbered from top to
bottom.

2.3.2.2

Digital Out

The following relay outputs are available for further signal processing. The maximum
rating for the relay contacts is 25V AC/1A.

Description of the operation mode of the Digital Out.

Trigger 1:
Is open in case Trigger Level 1 is exceeded or the machine is not in condition of
measuring.

Trigger 2:
Is open in case Trigger Level 2 is exceeded or the machine is not in condition of
measuring.

Zero active:
Is closed in case Zero function is running.

Ready:
Is closed in case machine is ready for measurement (Emission on, no error).

CAL active
Closed when machine is in calibrating routine.

CAL Request
Is opened in case of calibration request. During external calibration an open output
indicates that the external calibrated leak has to be closed.

Error
Open when an error is shown.

Warning
Open when a warning is shown.

Pin

Assignment

1

+24 V, constantly applied, power supply for
the INFICON partial flow valve or sniffer line.

2

GND24

3, 6

Input

4, 5, 7, 8

Output

Pin

Assignment

Pin

Assignment

1

+24V, bridged with pin 1
of socket „Digital In“

9

CAL request

2

GND_24V

10

Error

3

Trigger 1

11

Warning

4

Trigger 2

12

Purge

5

Reserved

13

Measure

6

Zero active

14

Recorder Strobe

7

Ready

15

Common dig. out

8

CAL active

16

Reserved