Default parameters

The following parameters are set like shown when in the menu of the UL5000 under
Settings

 Parameters saveload, „load default values“ is chosen.

Auto-scaling:

On

Scaling

logarithmic

Display range:

4 decades

Time axis:

32 seconds

LCD invers:

OFF

Background in stand by mode:

OFF

Automatic calibration request:

OFF

Mass:

4 (helium)

Recorder Output:

leak rate

Volume:

2

Zero time:

5 seconds

Leak rate unit:

mbar l/s

Mode:

Vacuum with
HYDRO•S

Trigger level 1:

1×10

-9

mbar l/s

Trigger level 2:

1×10

-8

mbar l/s

Leak rate external test leak (Vacuum):

1×10

-7

mbar l/s

Leak rate external test leak (Sniffer):

1×10

-5

mbar l/s

Vent delay:

2 seconds

Automatic purge:

OFF

Pressure unit:

mbar

Minimum volume:

0

Beep:

ON

Maximum evacuation time:

30 minutes

Audio Alarm Type:

Trigger Alarm

Maximum inlet pressure when sniffing:

1 mbar

Minimum Inlet pressure when sniffing:

0,1 mbar

Number of decimal place at leak rate displayed:

1

Scroll display:

On

Particle protection:

Off

Direct access to calibration:

On

Contamination protection:

Off

Switch off limit for contamination protection:

1E-3 mbar l/s

Control location:

Local

Alarm delay:

30 seconds

Leak rate filter:

I•Cal

Zero:

enable

HYDRO•S automatic

enabled