Hei-vap advantage, 4 hei-vap advantage – Heidolph Hei-VAP Advantage User Manual
Page 63

Errors and Troubleshooting
Hei-VAP Value, Hei-VAP Advantage
55
EN
8.4 Hei-VAP Advantage
Error / message
(error label)
Cause
Remedy
1
The transportation protection device is still
activated.
(only motor lift version)
Please raise the lift and
briefly switch the unit off
and then on again.
No data received at all for 3 se-
conds
Cable connection between the control
panel and the base unit was discon-
nected. An error occurred in the data
bus.
Restore the cable con-
nection between the con-
trol panel and the base
unit. Switch the power
switch off and on again.
Contact service
3
Heater is defective, i.e. the over-
heating protection was triggered.
If the medium was in the
heating bath: Let the
heating bath cool and re-
set the maximum tem-
perature limiter
If the medium was not in
the heating bath: Contact
service
Heating bath safety temperature
(+5 K) is exceeded
Contact service
Heating bath is completely evapo-
rated.
Deactivate the unit
Check if the maximum
temperature limiter has
responded. If yes, reset
(see chapter 9.1.1)
Fill in heat transfer fluid
Heating bath sensor variance
> 10 K
Contact service
Sensor monitoring via hardware
HB sensor broken or short circuit-
ed
Contact service
The temperature sensor for moni-
toring the heating bath control is
broken or the temperature of the
medium in the heating bath is
> 217 °C.
Cool down the heating
bath medium
Contact service
The temperature sensor for the
heating bath control is short cir-
cuited or the temperature is below
0 °C
Contact service
5
Motor is not rotating:
Motor overload shutdown
Deactivate the unit
Stop drive unit for cooling
down
Contact service
Table8-4: Troubleshooting table - Hei-VAP Advantage