Errors and troubleshooting, General errors, Errors and troubleshooting) – Heidolph Hei-VAP Advantage User Manual
Page 60: R 8, "errors and troubleshooting, 8 errors and troubleshooting, 1 general errors

Errors and Troubleshooting
52
Hei-VAP Value, Hei-VAP Advantage
EN
EN
EN
8 Errors and Troubleshooting
8.1 General errors
Error / message
Cause
Remedy
Unit cannot be switched on
Main plug not connected to the main pow-
er supply
Connect the main plug
with the power supply
Fuses defective or blown fuse
Replace fuses
(see chapter 9.2)
No heating function
Master switch is deactivated
Activate main switch
Fuses defective or blown fuse
Replace fuses
(see chapter 9.2)
Master switch is defective
Contact service
The connecting cable of the heating bath is
not connected.
Connect the connecting
cable of the heating bath
The heater of the heating bath is defective
Contact service
The maximum temperature limiter re-
sponded
If the medium was in the
heating bath: Let the
heating bath cool and re-
set the maximum tem-
perature limiter
If the medium was not in
the heating bath: Contact
service
Drive unit is not rotating
Master switch is defective
Contact service
Speed controller is on the left stop
(Hei-VAP value)
Turn the speed controller
to the right
Drive unit is defective
Contact service
Fuses defective or blown fuse
Replace fuses
(see chapter 9.2)
Motor lift is not functioning
Master switch is deactivated
Activate main switch
Master switch is defective
Contact service
Fuses defective or blown fuse
Replace fuses
(see chapter 9.2)
Lift is at end stop
Activate other arrow key
Mechanical system / motor is defective
Contact service
Height stop is not set accurately
Perform height adjust-
ment
(see chapter 4.5.7)