Additional conditions, 2 additional conditions – Heidolph Hei-VAP Advantage User Manual
Page 61

Errors and Troubleshooting
Hei-VAP Value, Hei-VAP Advantage
53
EN
Error / message
Cause
Remedy
No evacuation
Power switch Rotavac vario control or
Rotavac vario tec is not activated.
Vacuum valve is defective.
Activate the power switch
Fuses defective or blown fuse
Replace fuses
(see chapter 9.2)
Insufficient vacuum
System leaks
Check seals, tubing and
connections
Check stopcocks, grease
if necessary
Vacuum pump is defective
Observe manufacturer's
instructions of the vacu-
um pump
Unit suddenly switches off
(Hei-VAP Precision and
Hei-VAP Advantage)
Timer is programmed
Check and switch off the
timer, if necessary
(see chapter 7.4.1)
Fuses defective or blown fuse
Replace fuses
(see chapter 9.2)
Table 8-1: Troubleshooting table
8.2 Additional conditions
Additional conditions
Reaction
Vapor sensor is defective
The actual value curve is displayed “---“ instead of the vapor tem-
perature.
The vapor sensor is not con-
nected
The vapor temperature is not displayed in the actual value display
Heater can only be activated at
>0 °C (observe acetone, dry
ice)
Table 8-2: Additional conditions