Hei-vap value, 3 hei-vap value – Heidolph Hei-VAP Advantage User Manual
Page 62

Errors and Troubleshooting
54
Hei-VAP Value, Hei-VAP Advantage
EN
EN
EN
8.3 Hei-VAP Value
Error / message
(flashing code)
Cause
Remedy
**
No data received at all for 3 se-
conds
Cable connection between the control
panel and the base unit was discon-
nected. An error occurred in the data
bus.
Restore the cable con-
nection between the con-
trol panel and the base
unit. Switch the power
switch off and on again.
Contact service
***
Heater is defective, i.e. the over-
heating protection was triggered.
If the medium was in the
heating bath: Let the
heating bath cool and re-
set the maximum tem-
perature limiter
If the medium was not in
the heating bath: Contact
service
HB safety temperature (+5 K) is
exceeded
Contact service
Heating bath is completely evapo-
rated.
Deactivate the unit
Check if the maximum
temperature limiter has
responded. If yes, reset
(see chapter 9.1.1)
Fill in heat transfer fluid
Heating bath sensor variance
> 10 K
Contact service
Sensor monitoring via hardware
HB sensor broken or short circuit-
ed
Contact service
The temperature sensor for moni-
toring the heating bath control is
broken or the temperature of the
medium in the heating bath is
> 217 °C.
Cool down the heating
bath medium
Contact service
The temperature sensor for the
heating bath control is short cir-
cuited or the temperature is below
0 °C
Contact service
****
Potentiometer temp. is defective
(value)
Contact service
Potentiometer speed is defective
(value)
Contact service
*****
Motor is not rotating:
Motor overload shutdown
Deactivate the unit
Stop drive unit for cooling
down
Contact service
Table 8-3: Troubleshooting table - Hei-VAP Value