INSTALLATION
Reagent Line Assembly
1. Cut appropriate lengths of 1/16”ID Teflon tubing, not to exceed 5’ lengths. The ends need to be a
square-cut face.
2. Slide the flangeless nut over the tubing, with the nut threads facing the tubing end.
3. Slide the flangeless ferrule over the tubing, with the tapered portion of
the ferrule facing towards the nut. The large diameter of the ferrule
should be pushed to the end of the tubing.
4. AutoBlock reagent ports are located on the left side of AutoBlock
labeled #’s 1-10.
5. Insert the tubing with the flangeless ferrule into the receiving port
until it bottoms. While holding the tubing, tighten the Flangeless nut
finger tight. The tubing should not be able to move in the port.
Repeat for all ports available.
reference diagram for placement
Flangeless Ferrule
Flangeless nut
Placement of Reagent Rack
Position the six-bottle reagent rack (optional) on the left side of AutoBlock.
For dual systems, there may be two six-bottle reagent racks (optional).
AutoBlock Reagent Port
AutoBlock reagent ports are located on the left side of AutoBlock, labeled #’s 1-10.
The reagents have been set up in the following manner:
Single System
Dual System**
*Port #1
H20(1)
*Port #6
H20(6)
Port #2
HNO3(2)
Port #7
KMnO4(7)
Port #3
HCL(3)
Port #8
PERSULF(8)
Port #4
H2O2(4)
Port #9
NH2OH(9)
Port #5
H2SO4(5)
Port #10
H20(10)
*Ports #1 & #6 must remain H
2
0
**In a single system ports #6-10 have been plugged.
If these reagents do not suit your application, you may change them. Changing the reagents and ports
numbers are addressed in the SERVICE TAB – SETUP REAGENT section.
reagent plate
Vented Reagent Caps and Bottom of the Bottle Filter (optional)
Loosen the nut on the top of a vented reagent cap. Insert one end of reagent tubing through the nut until
it reaches the bottom of the 2.5L bottle. Hand tighten the nut.
A bottom of the bottle filter may be used to filer the reagent. Attach this to the end of the tubing before
inserting it into the 2.5L bottle.