SERVICE TAB
The SERVICE TAB is set up for AutoBlock to be calibrated,
maintained, and custom levels set for optimal performance.
service tab screen
Setup Reagents
Select SETUP REAGENTS button from the SERVICE TAB
Reagents are set to correspond with reagent ports. AutoBlock reagent ports are located on the
left side of AutoBlock, labeled #’s 1-10. The reagents have been set up in the following manner:
Single System
Dual System**
*Port #1
H20(1)
*Port #6
H20(6)
Port #2
HNO3(2)
Port #7
KMnO4(7)
Port #3
HCL(3)
Port #8
PERSUL(8)
Port #4
H2O2(4)
Port #9
NH2OH(9)
Port #5
H2SO4(5)
Port #10
H20(10)
*Ports #1 & #6 must remain H
2
O
**In a single system ports #6-10 have been plugged.
If the reagent arrangement or if other reagents are used, you may change the list.
In the following steps. Listed below are common reagents with viscosity ratings.
Reagent
Viscosity
H2O(1)
(ports 1 & 6, cannot be changed)
Low
HNO3(2) Nitric
Acid
Medium
HCL(3) Hydrochloric
Acid
Low
H2O2(4)
Hydrogen Peroxide 30%
Medium
H2SO4(5) Sulfuric
Acid
High
H2O(6)
(ports 1 & 6, cannot be changed)
Low
KMnO4(7) Potassium
Permanganate
Low
PERSUL(8) per sulfates
Low
NH2OH(9) Hydroxylamine
Low
H2O(10) Water
Low
Viscosity ratings control the pump speed to maximize efficiency and provide volume delivery accuracy.
Low Viscosity, fast pump speed, (reagent near the viscosity of water)
High Viscosity, slow pump speed