Environmental Express AutoBlock III User Manual
Page 17

MANUAL TAB
HOME
(blue section)
This operation will home the three lift motors and the injection arm motor.
The three lift motors will go to the down position and the injection arm will position itself over the drain.
Press the Home button
-A pop up message will appear: “HOME MOTOR?” Press Continue or Cancel button.
Cancel
You have canceled the command & return to the manual tab
Continue
The lift &injection arm motors will activate
A pop up message will appear: “RACKS ARE DOWN AND ARM IS LOCATED OVER THE
DRAIN.”, Press Continue button
FAN
(blue section)
By pressing this button the fan will turn on or off.
A confirmation message fan is on/off. Press the OK button.
The fan will automatically turn on when the block is over 40
0
C, and cannot be turned off below 40
0
C
Graphite
(pink section)
Range: 15
0
C-150
0
C
Set the temperature to the graphite block with the or
buttons, or by using the keyboard.
The temperature is set in
0
C, inc/dec by 0.1
0
C
After selecting a temperature press the
SET
button.
- A pop up message will appear: “THE BLOCK TEMPERATURE HAS BE SET” Press Continue button
The change will also reflect the readout on the upper right corner of the screen
NOTE:
The temperature of the graphite block will not be the temperature of the external thermocouple (sample
probe).
Test the sample temperature to achieve proper graphite temperature.
Heat the graphite block to 115
0
C, the sample probe should read 95
0
C
The use of the reflux cap (SC506) will raise the sample temperature 10
0
C
ENVIRONMENTAL EXPRESS
For technical support call 1-800-745-8218 or send email to [email protected]
5528MANUAL rev3
17
9/10/09