Environmental Express AutoBlock III User Manual
Page 18

MANUALTAB
Arm
(red section)
Select which column you would like a reagent injected to by the pull down menu
-Select DRAIN, 9,8,7,….1
-Press the MOVE button
-A pop up message will appear: “MOVE INJECTION ARM?”,
Press
Continue or Cancel button.
Cancel
You have canceled the command & return to the manual tab
Continue
The injection arm will move to the selected column
A pop up message will appear: “INJECTION ARM MOVE COMPLETE.”
Press Continue button
NOTE:
**Dual System**
There will be a left & a right splitter.
ARM (1-5) will inject thru the left splitter
ARM (6-10) will inject thru the right splitter
single unit system shown
dual unit has two splitters
Inject
(green section)
Range: 0.1ml-50.0ml
Select which column to be injected thru the ARM-section and move injection arm before proceeding.
The Inject command does not automatically purge or prep. Bring the injection arm to he drain and inject
6ml H20(1) to purge and 6ml of reagent to prep.
Amount
Select the amount of reagent to be injected with the
or
buttons or by using the keyboard.
The volume is set in ml. inc/dec by 0.1ml
Reagent
Select the Reagent to be injected from the drop-down menu. Reagents have been pre-installed.
Reference the SERVICE TAB-REAGENT SETUP for port configuration and to make changes
EXAMPLE: H20(1)
Reagent H20 is connected to port (1) on the AutoBlock unit
-Select REAGENT(PORT)
After selecting a reagent press the
INJECT
button
-A pop up message will appear: “INJECT REAGENT?” Press Continue or Cancel button.
Cancel
You have canceled the command & return to the manual tab
Continue
The reagent will inject the specified amount
A pop up message will appear: “REAGENT INJECTED.” Press Continue button
NOTE:
**Dual System**
REAGENTS (1-5) will inject thru the left splitter
REAGENTS (6-10) will inject thru the right splitter
ENVIRONMENTAL EXPRESS
For technical support call 1-800-745-8218 or send email to [email protected]
5528MANUAL rev3
18
9/10/09