6. To calibrate all inputs, repeat Steps 1 to 4 for each of the other input types (see

Table 9-2) until all three secondary analog input types have been successfully
calibrated.


Table 9-2 Secondary Analog Input Type Selection

Input Type No.

Input Type

Calibration Input

Link Jumper 10

Link Jumper 11

7

Remote Setpoint,
Linear DC mV

50 mV DC

Parked

Parked

8

Remote Setpoint,
Linear DC V

10 VDC

Parked

Fitted

9

Remote Setpoint,
Linear DC mA

20 mA DC

Fitted

Parked

NOTE: No calibration is required if the secondary analog input is to be used as a Remote Setpoint
Potentiometer Input or for Dual Setpoint switching.

9.4 Exit From Calibration Mode


To exit from Calibration Mode, press the DOWN and SCROLL keys simultaneously.

NOTE: An automatic exit is made from Calibration Mode if there is no key activity for
two minutes.

9.5 Calibration Check


1. Set the Controller to the required configuration (using link jumpers and front panel

entry) as described in the appropriate Site Manual.

2. Power-up the Controller and leave it powered-up for at least five minutes (for RTD

and DC linear inputs) or at least 30 minutes (for thermocouple inputs).

3. After the appropriate delay for stabilization has elapsed, check the calibration by

connecting the appropriate input source and checking a number of cardinal points.