CHAPTER TWO
QUICK STARTUP GUIDE
Turn the PVX on and off using the switch located on the bottom right corner of the
keypad. When PVX is initially turned on, a flash logo and blinking lights will be
displayed prior to entering into the main measurement screen. Note: This section is
primarily written as a basic startup guide only.
2.1 Selecting The Transducer Type
The first step in using the PVX is to select the transducer type stored in a setup.
There are 14 factory setups containing different transducer types that can be loaded,
modified and stored in a new setup location. B y selecting the transducer type from a
predefined setup list, the PVX can recall specific properties about the transducer,
setup the display area and predefined scope settings. Note: Once the transducer
has been selected, the PVX will store and recall this setup, containing the transducer
type, every time the PVX is powered on/off. This setup will only change if the user
physically selects another setup from the list. Therefore, if you have previously gone
through this section and selected the appropriate setup, proceed to the next section.
Use the following steps to select your setup (transducer type):
Selecting a Setup for use with a specific Transducer Type
1) Press the MULTI MODE key once to activate the load setup list containing
the transducer types and settings.
2) Use the UP and DOWN arrow keys to scroll through the setup list until the
target setup is highlighted.