Dakota Ultrasonics PVX User Manual

Page 42

background image

Dakota Ultrasonics

38

4) Press the ENTER key to display the Digits Edit Box.

5) Press the UP and DOWN arrow keys to scroll the highlighted value.

6) Press the LEFT and RIGHT arrow keys to scroll the digit locations.

7) Repeat steps 5 & 6 until the known thickness value is correctly displayed.

8) Press the OK key to calculate the velocity and return to the menu screen, or

ESC to cancel entering the velocity.

9) Finally, press the MEAS key to return to the measurement screen and begin

taking readings.

Note: CHECK YOUR CALIBRATION! Place the transducer back on the

calibration point. The thickness reading should now match the known

thickness. If the thickness is not correct, repeat the steps above.


Two Point Calibration

Note: Be sure that a probe zero has been done prior to performing this

calibration procedure, if a contact transducer in pulse-echo mode is being used.