PVX Precision Visual Thickness Gauge

105

Important Note: Remember to perform a probe zero function prior to

measuring. The default setup clears the probe zero value.

Important Note: If interface-echo or echo-echo modes are being used, be sure

to do a two -point calibration, which also sets the probe zero before measuring.

4)

Press the OK key to return to the measurement screen.