beautypg.com

APG DCR-1006 user manual User Manual

Page 20

background image

DCR-1005 and DCR-1006

Rev. A4, 10/08

20

Automation Products Group, Inc.

APG...Providing tailored solutions for measurement applications

Tel: 1/888/525-7300 • Fax: 1/435/753-7490 • www.apgsensors.com • [email protected]

Analog
The DCR-1006 has an Isolated 4-20 mA circuitry. To set up this feature, enter

the two end points. For performing a distance to level measurement, the zero
point of distance will be at the transducer and the 4-20 limits will be
programmed in the same units as mode 1. For volumetric monitoring, the zero
point will be at the empty point of the tank and the end points must be set in
the units used in the volume measurement. Keep this in mind when referring to
the diagram below. Remember that this circuit is Loop Powered. Power must be
supplied to the 4-20 mA terminal for this feature to operate.

MODE DESCRIPTION

PARAMETERS

EXPLANATION

24

4 mA distance

Units = mode 1

Sets the minimum value

Default = 3.00 ft.

of the analog output slope

25

20 mA distance

Units = mode 1

Sets the maximum value

Default = 4.00 ft.

of the analog output slope

26

4 mA calibration

Approx. 5000

Fine tunes the minimum
current sourced on the
analog output

27

20 mA calibration

Approx. 5000

Fine tunes the maximum
current sourced on the
analog output

External
P o w e r
Source

(-) (+)

4-20(+)

4-20(-)

Isolated 4 to 20 ma output

24V

DIG

GND

4-20 ma output

Analog Ground

Current sourcing 4 to 20 ma output

Analog Ground

4-20 ma output

4-20(+)

4-20(-)

This manual is related to the following products: