1s s – Eppendorf Excella E-24R Benchtop Incubator Shakers User Manual
Page 41
41
New Brunswick
Operating Manual
1
1
1
1
S
S
P
P
E
E
C
C
I
I
F
F
I
I
C
C
A
A
T
T
I
I
O
O
N
N
S
S
These specifications assume a maximum load of 34 pounds (15.5 kg), including platforms,
clamps, glassware and contents.
Excella E-24 & E-24R Incubator Shaker
SHAKING
Speed
1
Control Accuracy
Indication
Stroke/Orbit
50-400 rpm
1 rpm
Displayed in 1 rpm increments
¾-inch (1.9 cm)
TEMPERATURE
Range (E-24)
2
Range (E-24R)
2
Control Accuracy
Indication
Heater
7ºC above ambient temperature to 60°C
15ºC below ambient temperature (minimum 4ºC) to 60°C
0.1°C at 37°C
Displayed in 0.1°C increments
Long-life, low-watt density resistance-type heater with high temperature
thermostat
ALARMS
Visible and audible warning indication when speed deviates more than 5
rpm or temperature more than 1°C from setpoints, and when timer has
expired. Audible alarm can be muted.
LED DISPLAY
9/16-inch (1.4 cm)
RS-232
Data logging
SETPOINT RETENTION
All setpoints and operating status are retained in non-volatile memory
AUTOMATIC RESTART
Automatic restart after power is restored, indicated by flashing display
DRIVE
UniCentric™ counterbalanced drive with four permanently lubricated ball
bearings
DRIVE MOTOR
Solid-state brushless DC motor.
SAFETY
Drive Interrupt shuts off power to Shaker when lid opens.
Acceleration/deceleration circuit prevents sudden starts and stops,
minimizing both splashing and mechanical damage.
ELECTRICAL
REQUIREMENTS
100 Volts, 50/60 Hz
120 Volts, 60 Hz
230 Volts, 50 Hz
E-24: 800 VA per shaker
E-24R: 1500 VA per shaker
UL 61010A-1
CAN/CSA-C22.2 No. 1010.1
ETL REGULATORY
STANDARDS
UL 61010A-2-010
CAN/CSA-C22.2 No. 1010.2.010
CE REGULATORY
STANDARDS
See Declaration of Conformity, Section 11.1.
DIMENSIONS
Width
Depth (Front to Back)
Height
Height with lid open
22 inches (55.8 cm)
30 inches (76.2 cm)
24 1/16 inches (61.1 cm)
40⅛ inches (101.9 cm)
1
see first NOTE on following page
2
depending on ambient factors, such as relative humidity. Ambient temperature is measured one meter
from the front of the shaker
.