beautypg.com

1s s – Eppendorf Excella E-24R Benchtop Incubator Shakers User Manual

Page 41

background image

41

New Brunswick

Operating Manual

1

1

1

1

S

S

P

P

E

E

C

C

I

I

F

F

I

I

C

C

A

A

T

T

I

I

O

O

N

N

S

S


These specifications assume a maximum load of 34 pounds (15.5 kg), including platforms,
clamps, glassware and contents.

Excella E-24 & E-24R Incubator Shaker

SHAKING
Speed

1

Control Accuracy
Indication
Stroke/Orbit


50-400 rpm
 1 rpm
Displayed in 1 rpm increments
¾-inch (1.9 cm)

TEMPERATURE
Range (E-24)

2

Range (E-24R)

2

Control Accuracy
Indication
Heater


7ºC above ambient temperature to 60°C
15ºC below ambient temperature (minimum 4ºC) to 60°C
 0.1°C at 37°C
Displayed in 0.1°C increments
Long-life, low-watt density resistance-type heater with high temperature
thermostat

ALARMS

Visible and audible warning indication when speed deviates more than 5
rpm or temperature more than 1°C from setpoints, and when timer has
expired. Audible alarm can be muted.

LED DISPLAY

9/16-inch (1.4 cm)

RS-232

Data logging

SETPOINT RETENTION

All setpoints and operating status are retained in non-volatile memory

AUTOMATIC RESTART

Automatic restart after power is restored, indicated by flashing display

DRIVE

UniCentric™ counterbalanced drive with four permanently lubricated ball
bearings

DRIVE MOTOR

Solid-state brushless DC motor.

SAFETY

Drive Interrupt shuts off power to Shaker when lid opens.
Acceleration/deceleration circuit prevents sudden starts and stops,
minimizing both splashing and mechanical damage.

ELECTRICAL
REQUIREMENTS

100 Volts, 50/60 Hz
120 Volts, 60 Hz
230 Volts, 50 Hz

E-24: 800 VA per shaker
E-24R: 1500 VA per shaker

UL 61010A-1

CAN/CSA-C22.2 No. 1010.1

ETL REGULATORY
STANDARDS

UL 61010A-2-010

CAN/CSA-C22.2 No. 1010.2.010

CE REGULATORY
STANDARDS

See Declaration of Conformity, Section 11.1.

DIMENSIONS
Width
Depth (Front to Back)
Height
Height with lid open

22 inches (55.8 cm)
30 inches (76.2 cm)

24 1/16 inches (61.1 cm)

40⅛ inches (101.9 cm)

1

see first NOTE on following page

2

depending on ambient factors, such as relative humidity. Ambient temperature is measured one meter

from the front of the shaker

.




This manual is related to the following products: