Status Indicators


Four status indicator lights are located to the left of the

LED DISPLAY

. They are:

SET

Indicates that the shaker is in the

SET MODE

,

when setpoints are being displayed and can be
altered. This is activated by the

SELECT

key

or by pressing the

(UP)

or

(DOWN) KEY

.

POWER

Illuminates and blinks during power up or if
power is interrupted during a run. Press the

SELECT

key and change to another function to

turn off this indicator.

HEAT

Illuminates to indicate that the heater is on.

TIME

Indicates that the timer is in operation. The
shaker can be programmed to run for a preset
time from 0.1 to 99.9 hours. The timer can be
disengaged without stopping an ongoing run.

5.1.5 Function

Indicators


Four function indicator lights are located to the right of the

LED DISPLAY

. They indicate

the current parameter(s) being displayed:

C

Interior chamber temperature. Can be set from
4°C to 60°C, when in

SET MODE

, using the

(UP)

or

(DOWN) KEY

. It indexes at 0.1°C

increments unless the key is pressed for 4
seconds, after which it indexes more rapidly.

RPM

Revolutions per minute. When in

SET MODE

,

use the

(UP)

or

(DOWN) KEY

to change the

speed. It indexes at 1 RPM increments unless
the key is pressed for 4 seconds, after which it
indexes more rapidly.