Eppendorf Excella E-24R Benchtop Incubator Shakers User Manual

Page 21

background image

21

New Brunswick

Operating Manual

4. Press the

START/STOP KEY

again; the

TIME INDICATOR

will light and the shaker will

start the timed run.

If the shaker is stopped:


5. Press the

START/STOP KEY

. The shaker will start in untimed mode.

6. Press the

START/STOP KEY

again. The shaker will stop and the display will read

OFF

.

7. Press the

START/STOP KEY

a third time; the

TIME INDICATOR

will light and the

shaker will start the timed run.


To disable the visual alarm
(flashing

TIME INDICATOR

), press the

SELECT KEY

and

change to any other function.

To cancel the timer without stopping the shaker:

Repeat steps 1 and 2 above, then immediately press the

START/STOP KEY

. The

TIME

INDICATOR

will cease to flash and the display will read

OFF

.

6.6 Alarm

Functions

In addition to the visual alarm mentioned above, the Excella E-24/24R shakers have an
audible alarm that is activated at predetermined times, as explained in Section 1
(Overview). It can be deactivated by using the

MUTE

function:


1. Press the

SELECT

key until the

MUTE

indicator illuminates.

2. Press the

(UP)

or

(DOWN) KEY

to display

ON

, then press the

SELECT KEY

.


To reactivate the audible alarm:

1. Press the

SELECT

key until the

MUTE

indicator illuminates.

2. Press the

(UP)

or

(DOWN) KEY

to display

OFF

, then press the

SELECT KEY

.

6.7 Temperature

Setpoint

Press the

SELECT KEY

until the function

C INDICATOR

illuminates. The temperature can

be set from 5

C above ambient temperature to 60C (non-refrigerated units) or from 4C

to 60

C (refrigerated units). Increasing or decreasing the setpoint is accomplished with

the

(UP)

or

(DOWN) KEY

.


During operation, if the temperature of the chamber is more than 1.0

C higher or lower

than the temperature setpoint, an alarm is triggered. This alarm consists of a flashing

C