beautypg.com

Eppendorf Excella E-24R Benchtop Incubator Shakers User Manual

Page 23

background image

23

New Brunswick

Operating Manual

6.9 Power

Failure

In the event of a power failure, the Excella E-24/24R shakers are equipped with an
automatic restart function.

If the shaker was in operation prior to the power interruption, when power is restored the
shaker will begin to operate at its last entered setpoint. The

LED DISPLAY

will flash and

the audible alarm will sound, indicating that a power failure has occurred. Press any key
to cease the flashing in the display and the audible alarm.

6.10 Speed

Calibration

To calibrate the shaking speed:

1. Set the shaker to a speed that can easily be measured. If you are using a strobe,

minimum speed should be 250 RPM.

2. Compare the reading on the display to the measured reading.

If an adjustment is needed:

1. Press the

SELECT KEY

until the

RPM

indicator light illuminates.

2. Press the

(UP)

and

(DOWN) KEYS

simultaneously. The display will indicate

CAL

.

3. Press either the

(UP)

or

(DOWN) KEY

to change the displayed value to match the

measured speed.

4. Press the

(UP)

and

(DOWN) KEYS

simultaneously to save the adjustment.

5. Turn unit

OFF

using the power switch, then turn it back

ON

.

This manual is related to the following products: