Caution – Eppendorf Excella E-24R Benchtop Incubator Shakers User Manual
Page 20

20
Excella E-24/24R Incubator Shakers
M1352-0050
Operating Manual
6.3
Continuous (Untimed) Run
1. Press
SELECT
until the
RPM INDICATOR
is illuminated.
2. If the display indicates that the shaker is
OFF
, press the
START/STOP KEY
.
3. Press either the
(UP)
or
(DOWN) KEY
to enter
SET MODE
(the
SET INDICATOR
will
illuminate).
4. Set the speed by using the
(UP)
or
(DOWN) KEY
until the desired setpoint is
displayed. Continued pressure the
(UP)
or
(DOWN) KEY
will cause the setting to
change more rapidly.
NOTE:
The setpoint may be changed during a run without stopping the shaker by
following steps 2-4 above. During speed changes, a visual alarm (flashing
RPM INDICATOR
) will flash and an audible alarm will sound until the speed
returns to within 5 rpm of the setpoint.
6.4
Checking Any Setpoint
1.
Press
SELECT
until the desired indicator is illuminated.
2.
Briefly press either the
(UP)
or
(DOWN) KEY
to enter the
SET MODE
and display
the current setpoint.
CAUTION!
Holding the the
(UP)
or
(DOWN) KEY
for more than 0.5 seconds causes the
speed setpoint to change. Should this occur, resetting will be necessary.
6.5 Timed
Functions
The shaker may be programmed to automatically stop after a preset time period of 0.1 to
99.9 hours. There must be power to the shaker in order to set the timer, although a timed
run can be initiated while the unit is either stopped or operating.
To set the timer:
1. Press the
SELECT KEY
until the
HRS INDICATOR
is illuminated.
2. Press either the
(UP)
or
(DOWN) KEY
to enter the
SET MODE
and set the desired
run time, between 0.1 and 99.9 hours.
If the shaker is stopped, skip to Step 5 below. If the shaker is already running:
3. Press the
START/STOP KEY
. The shaker will stop and the display will read
OFF
.