Symptom(s)

Probable Cause(s) & Solution(s)

Heater fuse blown: replace.
Compressor fuse blown: replace.
Compressor over-pressure switch activated: call for service.
Ambient temperature too high or too low: cool or heat the
room as needed.
Defective heater: call for service.
Defective refrigeration system: call for service.

Incubator does not
reach set temperature.

Incorrect temperature indication (see below).
Defective RTD assembly: call for service.

Incorrect temperature
indication.

Defective main board: call for service.

8.2 Product

Returns

Should you need to return your Excella E-24/24R to Eppendorf for any reason, first
contact Customer Service to obtain a Returned Material Authorization (RMA) number.
This number must appear on the outside of the shipping container, otherwise Eppendorf
Receiving will refuse to accept the shipment.

In addition, you must also certify that the instrument being returned has been thoroughly
cleaned and decontaminated. A form for this purpose is provided in Section 12 of this
manual; it can also be downloaded from our website (

www.nbsc.com

). A copy of this

completed Return Authorization and Decontamination Certificate must be attached to the
outside of the container, with a second copy packed inside with the instrument.

8.3

Opening the Service Compartment

WARNING!

Before opening the Service Compartment, always turn off the shaker and
disconnect the power cord from the power supply.

The Service Compartment contains the shaker’s electronic and temperature control
components. Normally, this compartment should be accessed by authorized service
technicians only.
You may, from time to time, need to remove the access panel in order
to replace fuses (see Section 8.4 below).