2 cleaning, Eppendorf piezoxpert™ — operating manual – Eppendorf PiezoXpert User Manual
Page 24
Eppendorf PiezoXpert™ — Operating manual
24
7.2 Cleaning
Moisten a towel with mild cleaning fluid and dimineralized water and remove contamination
from the outside of the device and actuator.
2. Press the left and right plastic springs 1
together. Completely remove fuse holder 2.
3. Replace the defective fuses. Reinsert the
fuse holder.
1
2
DANGER!
DANGER! Electric shock as a result of penetration of liquid.
Switch off the device and disconnect it from the power supply before starting
cleaning or disinfecting work.
Do not allow any liquids to penetrate the inside of the housing.
Do not spray clean/spray disinfect the housing.
Only connect the device to the power supply if it is fully dried inside and out.
NOTICE!
NOTICE! Damage from the use of aggressive chemicals.
Do not use any aggressive chemicals on the device or its accessories, such as
strong and weak bases, strong acids, acetone, formaldehyde, halogenated
hydrocarbons or phenol.
If the device becomes contaminated with aggressive chemicals, clean it
immediately with a mild cleaning agent.
NOTICE!
NOTICE! Damage to the actuator due to penetrating liquid.
Penetrating liquid can damage the piezo element.
Do not clean the actuator under running water.
EN
Operating manual