3 optimizing parameters, Eppendorf piezoxpert™ — operating manual – Eppendorf PiezoXpert User Manual

Page 20

background image

Eppendorf PiezoXpert™ — Operating manual

20

1. Mount the capillary in the actuator grip head (see Mounting the capillary on p. 14).
2. Using the CellTram vario, fill the mounted capillary with oil from the rear.
3. Absorb the medium via the front capillary opening.
4. Initially, select a higher setting (e.g., 20) for the Int. (intensity) parameter.
Alternative capillary filling

Weigh down the capillary by filling it with liquids with higher specific densities.

Fill the capillary with a liquid with a high specific density, free of air bubbles, and then with oil.

In this case, reduce all parameters, especially Int. (intensity).

5.4.2

Preparing capillaries with smaller opening diameters

The following section contains a description of how to prepare capillaries with smaller opening

diameters, e.g., for mouse ICSI.

1. Fill the capillary from the rear using a microloader, e.g., Fluorinert™ FC-77.
2. Mount the capillary in the actuator grip head (see Mounting the capillary on p. 14).
3. Absorb the medium via the front capillary opening.
4. Initially, select a lower setting (e.g., 10) for the Int. (intensity) parameter.

5.4.3

Optimizing parameters

The PiezoXpert offers you a wide range of setting options. To determine the optimal parameters

for your application, proceed as follows:

1. Set the parameters for the speed and quantity of the piezo impulses to 1.

Speed: 1
Pulse: 1

2. Set the parameter for Int. (intensity) to a low value (e.g., 10).

Int.: 10 (example)

3. Incrementally increase the value for Int. until the piezo impulses is strong enough to penetrate

the cell wall.

4. Now, fine tune the Speed and Pulse parameters.
Or, you can start with a high value for the intensity (e.g., Int.: 30) and optimize the strength of the

pulse as the values decrease.
If you do not have success with the optimized parameters, you have to take into consideration

that the PiezoXpert cannot deliver the required contribution on its own.

WARNING!

WARNING! Damage to health due to toxic, radioactive or aggressive chemicals.

Observe the national regulations for handling these substances as well as the

material safety data sheets and manufacturer's application notes.

Wear personal protective equipment (PPE).