2 error messages, 7 maintenance, 1 replacing fuses – Eppendorf PiezoXpert User Manual

Page 23: Maintenance 7.1 replacing fuses, Eppendorf piezoxpert™ — operating manual

background image

Eppendorf PiezoXpert™ — Operating manual

23

6.2 Error messages

7

Maintenance

7 Maintenance

7.1 Replacing fuses

The fuse holder is located between the power connection socket and the power switch. The fuses

may only be replaced with the same type of fuses.

Hint!

If the recommended troubleshooting measures are not successful after several

attempts, please contact your local Eppendorf partner. The addresses can be found on

the website: www.eppendorf.com. The addresses of Eppendorf subsidiaries can be

found on the second to last page of this operating manual.

Symptom/

message

Cause

Remedy

Display remains

dark.

The device was not connected

to the power supply or the

power switch was not switched

on.

Check the power connection

and power cable.

Switch on the device.

The display shows:

The actuator is not connected.

Error messages can be

acknowledged using any key.

Connect the actuator.

Check the actuator connection

to the PiezoXpert.

The actuator is defective.

Error messages can be

acknowledged using any key.

Replace the actuator.

Have the defective actuator

inspected by Eppendorf.

An error message

appears in the

display.

A control function has detected

an error.

Error messages can be

acknowledged using any key.

Switch the device off and back

on again.

Contact Technical Service if

error messages continue to

appear.

A B

Please connect

Actuator!

DANGER!

DANGER! Electric shock.

Switch off the device and disconnect the power plug before starting maintenance or

cleaning work.

1. Disconnect the power plug.

EN

Operating manual