Error codes – IKA RV 10 control FLEX auto User Manual
Page 68

68
Any malfunctions during operation will be identified by an error message on the display.
Once a serious error message has been displayed, the lift moves to the top end position and the device can no longer be operated. The lift can be
operated again.
Proceed as follows in such cases:
Switch off device using the main switch
Carry out corrective measures
Restart device
Error
Effect
Cause
Correction
Communication Distillation cancelled in
IR interface obstructed
- Press ESC to confirm
error bath
"Auto-distillation" mode
Heating bath switched off
- Switch on heating bath
- Check and clean IR interface
No rotation
Distillation stopped,
Rotary drive does not start or
- Switch device off and on at mains switch
no rotary drive
exhibits very large speed variation
No temperature Distillation cancelled in
No differential temperature increase - Press ESC to confirm
difference
"Auto-distillation" mode
of cooling water
- Check flow of cooling water
- Check that the cooling water hoses are fitted correctly
Temperature
Distillation stopped
Differential temperature outside
- Press ESC to confirm
out of range
measurement range (> 7 K)
- Check that the cooling water hoses are fitted correctly
No vacuum
Distillation stopped
Vacuum target value not reached
- Press ESC to confirm
Vacuum cannot be controlled
- Check that the glassware and hose connections
are properly sealed
- Switch on the pump
- Check the solenoid valve connections
No venting
Vacuum cannot be vented
No venting of condenser
- Switch off the device
- Vacuum sensor faulty
- Manually vent the glassware using the stop
valve
Water valve
Water runs continuously into
Water valve not closed
- Press ESC to confirm
not closed
cooling circuit
- Check the connection
Flow rate
Distillation cancelled in
Flow rate of cooling water out of range - Press ESC to confirm
out of range
"Auto-distillation" mode
- Readjust flow rate
- Switch off flow rate control (for dry ice coolers)
Condenser
Distillation stopped
Condenser overload
- Press ESC to confirm
overload
Distillate may be able to enter the
- Change process parameters, e.g.
vacuum pump
- Increase pressure
- Reduce bath temperature
- Increase condenser flow rate within specified range
Distillation
Distillation stopped
The system detected that there was
- Press ESC to confirm
dry
no more medium in the evaporating - Alternatively, activate the “Drying” option
flask during the distillation.
in the “Settings” menu
Lift not in
Distillation stopped
Obstruction in path of lift
- Press ESC to confirm
end position
Lift motor switches off
- Check the path of the lift.
- Remove any obstructions.
Calibration error Calibration reset to factory setting
Out of calibration range
- Press ESC to confirm
(max. +/-0.5 K)
- Check that the sensor is connected correctly
- Carry out a new calibration procedure
Memory error
Calibration parameter for vacuum reset
Error by checking contents of memory Vacuum must be recalibrated
to factory setting
Vacuum leak
Distillation stopped
Leak in vacuum system
Press ESC to confirm
Locate leak
Error codes