beautypg.com

BNC 7000 Series Programmer Manual User Manual

Page 7

background image

7

The TCP/IP setup requires the following:

IP Address: Every computer/electronic device in a TCP/IP network requires an IP address. An IP

address has four numbers (each between 0 and 255) separated by periods.

For example: 192.168.1.50 is a valid IP address.

Subnet Mask: The subnet mask distinguishes the portion of the IP address that is the network ID

from the portion that is the station ID. The subnet mask 255.255.0.0, when applied to the IP address

given above, would identify the network ID as 192.168 and the station ID as 1.50. All stations in the

same local area network should have the same network ID, but different station IDs.

Default Gateway: A TCP/IP network can have a gateway to communicate beyond the LAN identified

by the network ID. A gateway is a computer or electronic device that is connected to two different

networks and can move TCP/IP data from one network to the other. A single LAN that is not

connected to other LANs requires a default gateway setting of 0.0.0.0. If you have a gateway, then the

default gateway would be set to the appropriate value of your gateway.

MAC Address: A MAC address is a unique 48-bit value that identifies a network interface card to the

rest of the network. Every network card has a unique MAC address permanently stored into its

memory.

Interface between the instrument and other devices on the network is via a category five (CAT-5)

interface cable connected to a network. This cable uses four twisted pairs of copper insulators

terminated into an RJ45 connector. CAT-5 cabling is capable of supporting frequencies up to 100 MHz

and data transfer speeds up to 1 Gbps, which accommodates 1000Base-T, 100Base-T, and 10Base-T

networks.

The instrument can be remotely programmed using the VXI-11 protocol. A VISA I/O library (like NI-

VISA™) is used on the server side to facilitate the communications. A VISA installation on the

controller is a prerequisite for remote control over VXI-11 interface. VISA is a standardized software

interface library providing input and output functions to communicate with instruments. For more

information about VISA refer to the VISA library supplier’s documentation.

The SCPI command set listed in the SERIES 7000 programmer’s manual applies to LAN programming

as well.

Only the IP address or the device name is required for link setup. The IP address/device name is part

of the "visa resource string" used by the programs for identification and control of the instrument. The

visa resource string has the form:

TCPIP::ipaddr::inst0::INSTR

ipaddr has to be replaced by the IP address or the computer name of the instrument.

For instance, if the instrument has the IP address 192.168.1.50, TCPIP::192.168.1.50::inst0::INSTR is

the valid resource name. Specification of inst0 in the resource name is optional. In this example, also

Berkeley Nucleonics Corporation 2955 Kerner Blvd., San Rafael, CA 94901

Phone: 415-453-9955, Fax: 415-453-9956, Email: [email protected], Web: www.berkeleynucleonics.com