2.8 White Balance

This procedure is to adjust the color balance.

1

Remove the “White Paper” and then place the “Setup Adjustment Sheet” on
the Adjustment Plate.

NOTE !

Refer to “2.4 CCD Gain” above to place the “Setup Adjustment
Sheet”. If the glass plate is attached, close it.

2

Select “7. White Balance” on the pull-down list under “Detail & Adjustment”,
and then click “Run”.

3

Click “OK” on the confirmation screen to confirm the test chart.

4

The window indicating that adjustment is underway will appear. When the
window disappears, the adjustment is completed. Proceed to the next
adjustment, “2.9 Tone Curve”.