Calibration commands (cont’d) – Agilent Technologies 75000 Series C User Manual
Page 87

Calibration Commands (cont’d)
•
CALibration:DATA:FILTer
constants that are used to determine the frequency points that will be
calibrated for the 10 MHz filter. The query form returns the current
constants in IEEE-488.2 definite block data format. See the AC
Flatness Adjustment procedure for the 10 MHz filter for more
information on the use of this command.
•
CALibration:DATA:SKEW transfers the calibration
constant that is used by the skew DAC to synchronize the AFG’s
DAC’s. The query form returns the current constant in IEEE-488.2
definite block data format. See the Skew DAC Adjustment procedure
for more information on the use of this command.
NOTE
The CAL:DATA:FILT and CAL:DATA:SKEW commands are available only
on units with firmware rev A.02.00 or higher (use the *IDN? command to
determine the AFG’s firmware revision).
•
CALibration:STATe:AC
corrections using the calibration constants. If
corrections will be used. If
used.
•
CALibration:STATe:DC
corrections using the calibration constants. If
corrections will be used. If
used.
•
CALibration:STATe
corrections using the calibration constants. If
corrections will be used. If
used.
•
*PUD stores the specified data in non-volatile "protected
user data" memory. The data must be sent in IEEE-488.2 definite or
indefinite block format. The query form (*PUD?) returns the
current protected user data in IEEE-488.2 definite block format.
Agilent E1445A Service Manual
Adjustments 85