Agilent Technologies 75000 Series C User Manual
Page 52

Test 2-5: AC Flatness - 250 kHz Filter (cont’d)
Test Procedure
(cont’d)
3.
Set the AFG output to the reference frequency (1 kHz):
FREQ 1000
Set frequency
4.
Measure the amplitude with the DMM and convert the reading to
dBm. Note the result for use in step 6:
Reference Level
(
dBm
)
=
20
×
log
ï
Reading
(
volts
)ï+
13.0103
Perform steps 5 - 6 for each frequency listed in Table 2-5:
5.
Set the AFG output:
FREQ
Set frequency
where
6.
Measure the amplitude with the DMM and convert the reading to
dBm. Calculate and record the error relative to the reference level
calculated in step 4:
Reading
(
dBm
)
=
20
×
log
ï
Reading
(
volts
)ï+
13.0103
Error
(
dB
)
=
Reading
(
dBm
)
−
Reference Level
(
dBm
)
50 Verification Tests
Agilent E1445A Service Manual