Test 2-5: AC Flatness - 250 kHz Filter
Description
The purpose of this test is to verify that the AFG meets its specifications for
AC flatness with the 250 kHz filter enabled.
Equipment Setup
•
Connect equipment as shown in Figure 2-6
•
Set DMM to ACV, autorange
Test Procedure
1.
Reset the AFG:
*RST;*CLS
Reset AFG and clear
status registers
2.
Set up the AFG to output a 24 dBm sinewave with the 250 kHz filter
enabled:
VOLT 24DBM;
Set amplitude
:OUTP:LOAD 50
50 ohm load
OUTP:FILT:FREQ 250KHZ
250 kHz filter
OUTP:FILT ON
Enable filter
INIT:IMM
Initiate waveform
Figure 2-6. Equipment Setup for Test 2-5 and Test 2-6
Agilent E1445A Service Manual
Verification Tests 49