Test 2-4: AC Accuracy (cont’d)
Test Procedure
(cont’d)
4.
Set the AFG output amplitude:
VOLT
Set amplitude
where
5.
Trigger the DMM and record the reading.
Table 2-4. AC Accuracy Test Points
Amplitude
(volts rms)
Filter
Test
Limits
±
(dB)
7.2390
6.4500
5.7500
4.5660
2.8818
1.4444
0.2290
7.2390
7.2390
None
None
None
None
None
None
None
250 kHz
10 MHz
0.10
0.15
0.15
0.15
0.15
0.15
0.15
0.10
0.10