Functional Verification: Start Arm In Test (cont’d)
Test Procedure
(cont’d)
4.
Set up the AFG to output a 1 MHz sinewave, with an external Start
Arm source:
FREQ 1E6;
Set freq to 1 MHz
:VOLT 4VPP
Set AFG amplitude
ARM:LAY2:SOUR EXT
External Start Arm
source
INIT:IMM
Initiate
5.
Verify that no signal appears on the scope. Send the following
command to the Command Module to provide a Start Arm signal to
the AFG:
OUTP:EXT:LEV 0
6.
Verify that a 1 MHz sinewave appears on the scope.
22 Verification Tests
Agilent E1445A Service Manual