CALibration:SECure[:STATe] ]
(
The security code is required for CAL:SEC:STAT OFF, but the code
is optional for CAL:SEC:STAT ON. The *RST command also
enables calibration security.
•
CALibration[:DC]:BEGin starts the DC calibration sequence and
sets up the AFG for the first calibration point.
•
CALibration[:DC]:POINt?
that the appropriate calibration constant(s) can be calculated. The
AFG returns two numbers: (1) the current calibration point, and (2)
an error code (see Appendix B of the Agilent E1445A User’s Manual
for more information about AFG errors). Any non-zero error code
indicates a failure. This command also sets up the AFG for the next
calibration point.
•
CALibration:DATA[:DC]
constants to the AFG. The DC calibration procedure described in
this chapter should be used in place of this command. The query
form returns the current DC constants in IEEE-488.2 definite block
data format.
•
CALibration:DATA:AC[1]
constants that are used with the 250 kHz filter. The query form
returns the current constants in IEEE-488.2 definite block data
format. See the AC Flatness Adjustment procedures for more
information on the use of this command.
•
CALibration:DATA:AC2
constants that are used with the 10 MHz filter. The query form
returns the current constants in IEEE-488.2 definite block data
format. See the AC Flatness Adjustment procedures for more
information on the use of this command.
84 Adjustments
Agilent E1445A Service Manual