Test 2-6: AC Flatness - 10 MHz Filter (cont’d)
Test Procedure
(cont’d)
7.
Set up the Power Meter:
Units - Watts
Power Range - auto
Reference Oscillator - ON
NOTE
Follow the Power Meter manufacturer’s instructions for performing an
autocalibration and correcting for the power sensor.
8.
Connect the equipment as shown in Figure 2-7:
9.
Set the Power Meter expected frequency to the crossover frequency
(100 kHz). Measure the AFG output power and convert the reading
to volts:
Reading
(
volts
)
=(
√
ï
Reading
(
watts
)
п
Ч
50
)
Figure 2-7. Equipment Setup for Test 2-6
Agilent E1445A Service Manual
Verification Tests 53