Test 2-9: total harmonic distortion (cont’d) – Agilent Technologies 75000 Series C User Manual
Page 67

Test 2-9: Total Harmonic Distortion (cont’d)
Test Procedure
1.
Reset the AFG:
*RST;*CLS
Reset AFG and clear
status registers
2.
Set the AFG to output a sinewave with the 10 MHz filter enabled:
VOLT 24DBM
Set AFG amplitude
OUTP:FILT:FREQ 10 MHZ
Set filter to 10 MHz
OUTP:FILT ON
Enable filter
INIT:IMM
Initiate waveform
Perform steps 3 - 6 for each frequency listed in Table 2-9:
3.
Set AFG output frequency:
FREQ
Set frequency
where
4.
Set the Spectrum Analyzer center frequency to the output frequency
of the AFG. Measure the peak amplitude of the fundamental in dBm.
Note the result for use in step 6.
5.
Set the Spectrum Analyzer center frequency to the second harmonic (
2
× fundamental frequency ). Measure the peak amplitude of the
second harmonic in dBm.
6.
Repeat step 5 for third through ninth harmonics. Calculate total
harmonic distortion as shown below:
thd
(
dBc
)
=
20
⋅
log
ж
и
√
result
2
2
+
result
3
2
+
…
+
result
9
2
ц
ш
where
result
n
(
volts
)
=
10
[
(
n
th
Harmonic
(
dBm
)
−
Fundamental
(
dBm
))
/20 ]
Agilent E1445A Service Manual
Verification Tests 65