20. ERROR COMMANDS
Err2
- Value beyond the zero range
Err3
- Value beyond the tare range
Err4
- Calibration mass or start mass beyond the acceptable
range (
±1% for weight, ±10 for start mass)
Err5
- Mass of a single piece lower than the scale division
Err8
- Exceeded the time for tarring, zeroing, start mass
adjustment or span adjustment
NULL
- Zero value from the AD converter
FULL2
- Measurement
range overflow
LH
- Start mass error, the mass on the weighing platform is
beyond the acceptable range (-5% to +15% of start mass)
5-FULL
- Display range overflow in totalizing
Notice:
1. Errors: Err2, Err3, Err4, Err5, Err8, null, that appear on the display
are also signalled by a short beep sound (about 1 sec.);
2. Error FULL2 that appears on the display is also signalled by
a continuous sound until the cause of error disappears.