RADWAG Low profiled scales User Manual
Page 58

• Put a weight of the displayed mass value on the pan and press
.
The calibration process will start which is signalled by the message:
• After completion of the process of calibration the following screen
will appear
• Take off the weight , then the following sequence of screens
will appear
• Calibration process can be terminated anytime by pressing
which is signalled by the following message on the display:
• Return to weighing with saving changes that have been made.
Caution:
If the calibration process (span adjustment) lasts longer than 15 the
emitted. Press
to perform calibration again with more stable
ambient conditions!
- 58 -