6 Commissoning, operation
26
WIKA operating instructions digital indicator DI10
6.5.2 Additional setpoints
Menu level Parameter-Ebene
SPCt
Setpoints - Number of additional setpoints, SPCt:
0
In addition to the start and end value, 8 extra setpoints can be
defined to linearize non-linear sensor values. Only the
activated setpoint parameters are displayed.
dIS1
Display values for setpoints dIS1 … dIS8:
8
8
8
8
nOCA
CAL
Under this parameter the value of the setpoints is defined.
With sensor calibration, as with end value/offset, you will be
asked at the end whether a calibration should be made.
InP1
Analogue values for setpoints INP1 … INP8:
0
0
0
0
The setpoints are only displayed with the works calibration (4-
20 mA). Here, the desired analogue values can be freely
selected. The input of constantly rising analogue values must
be carried out by the customer/user.