WIKA DI10 User Manual
Page 24
6 Commissoning, operation
24
WIKA operating instructions digital indicator DI10
Menu level Parameterization level
LI-2
Limit value /Limits, LI-2:
0
0
0
0
For both limit values, two different values can be
parameterized. With this, the parameters for each limit value
are called up one after the other.
HY-2
Hysteresis for limit values, HY-2:
0
0
0
0
For both limit values, a hysteresis function exists that reacts
according to the functional principle (operating current /
quiescent current).
Fu-2
Function if display falls below / exceeds limit value, FU-2:
HIgH
Louu
(Low)
To indicate if the value falls below the lower limit value, Louu
can be selected (LOW = lower limit value) and if it goes above
the upper limit value, high can be selected (HIGH = upper limit
value). LOW corresponds to the quiescent current principle
and HIGH to the operating current principle.