6 Commissoning, operation
18
WIKA operating instructions digital indicator DI10
6.4 Setting up the device
Switching on
Once the installation is complete, you can start the device by applying the
current loop. Check beforehand once again that all the electrical connections
are correct.
Starting sequence
For 1 second during the switching-on process, the segment test (8 8 8 8) is
displayed, followed by an indication of the software type and, after that, also
for 1 second, the software version. After the start-up sequence, the device
switches to operation/display mode.
Standard parameterization
To be able to parameterize the display, press the [P] key in operating mode
for 1 second. The display then changes to the menu level with the first menu
item tYPE.
Menu level Parameterization level
tYPE
Selection of the input signal, tYPE:
4-20
SEnS
There are two measuring input options for the current loop: 4-
20 mA as works calibration (without application of the sensor
signal) and SEnS as sensor calibration (with the sensor
applied). Confirm the selection with [P] and the display
switches back to the menu level.