6 Commissoning, operation
WIKA operating instructions digital indicator model DI10
19
Menu level Parameterization level
End
Setting the measuring range end value, END:
8
8
8
8
nOCA
CAL
Set the end value from the smallest to the highest digit with
[
▲
] [
▼
] and confirm each digit with [P]. A minus sign can only
be parametrized on the highest value digit. After the last digit,
the display switches back to the menu level.
If SEnS was selected as the input option, you can only select
between nOCA and CAL. With nOCA, only the previously set
display value is taken over, and with CAL, the device takes
over both the display value and the analogue input value.
OFFS
Setting the measuring range start/offset value, offs:
8
8
8
8
nOCA
CAL
Enter the start/offset value from the smallest to the highest
digit [
▲
] [
▼
] and confirm each digit with [P]. After the last digit
the display switches back to the menu level.
If SEnS was selected as the input option, you can only select
between nOCA and CAL. With nOCA, only the previously set
display value is taken over, and with CAL, the device takes
over both the display value and the analogue input value.