WIKA DI10 User Manual
Page 20
6 Commissoning, operation
20
WIKA operating instructions digital indicator DI10
Menu level Parameterization level
dot
Setting the decimal point, dot:
0
0.0
0.00
0.000
The decimal point on the display can be moved with [
▲
] [
▼
]
and confirmed with [P]. The display then switches back to the
menu level again.
SEC
Setting the display time, SEC:
0.1
0.9 dann 1.0
10.0
The display time is set with [
▲
] [
▼
]. The display moves up in
increments of 0.1 up to 1 second and in increments of 1.0 to
10.0 seconds. Confirm the selection by pressing the [P]
button. The display then switches back to the menu level
again.
run
Activation / deactivation of the programming lock and
completion of the standard parameterization, run:
ULOC
LOC
With the aid of the [
▲
] [
▼
] keys, you can choose between the
deactivated key lock ULOC (works setting) and the activated
key lock LOC. Make the selection with [P]. After this, the
display confirms the settings with "- - - -", and automatically
switches to operating mode.
If LOC was selected, the keyboard is locked. To get back into
the menu level, you must press [P] for 3 seconds in operating
mode. You must now enter the CODE (works setting 1 2 3 4)
that appears using the [
▲
] [
▼
] keys plus [P] to unlock the
keyboard. FAIL appears if the input is wrong.