WIKA operating instructions models TR10-L, TC10-L (Ex d)

GB

WARNING!

Before storing the instrument (following operation), remove any residual media.

This is of particular importance if the medium is hazardous to health, e.g. caustic,

toxic, carcinogenic, radioactive, etc.

6. Commissioning, operation

6.1 Removal and installation of the measuring insert

When servicing is necessary, the flame path fitting must be renewed when replacing the

measuring insert. When recalibrating, make sure that both surfaces of the flameproof joint

(flame path fitting and measuring insert) are not damaged when removing the measuring

insert.

6.2 Electrical connection
Connection to terminal block

For the electrical specifications (e.g. connection diagrams, tolerance values, etc.) please

refer to the data sheets TE 60.12 (for TR10-L) and TE 65.12 (for TC10-L).

Connection to built-in transmitter

For the electrical specifications (e.g. connection diagrams, tolerance values, etc.) please

refer to the relevant operating instructions and/or data sheet for the built-in head-mounted

transmitter.

Junction between Ex d cable gland and connection head

Threads M20 x 1.5: tightening torques 12 Nm

Threads ½ NPT: tightening torques 30 Nm

Junction between cable and Ex d cable gland

Screw the male nut tightly into the adapter (use appropriate tools!)

During installation, take care to

Avoid distorting the cable sheath when tightening the male nut.

Avoid cutting too deep into the cable sheath.

Use suitable cable.

Be careful of the clamping zone of the cable gland.

6. Commissioning, operation