beautypg.com

Myron L CTCIITD User Manual

Page 28

background image

C.

STANDARD SOLUTIONS & BUFFERS

1. CONDUCTIVITY/TDS

STANDARD SOLUTIONS

Your CTCII Chemical Treatment Controller has been factory

certified to meet certain specifications based on the appropriate

Myron L Company NIST traceable standard solution. This same

solution is available to you.

442-3000

3900 µS/µM Conductivity and 3000 ppm TDS

Other standard solutions are available to meet your needs or

requirements.

DESCRIPTION

The Myron L Company manufactures three basic types of

conductivity/TDS standard solutions — KCl, NaCl and 442™. See

below.

Most Myron L standard solution bottles show three values

referenced at 25°C:

Conductivity in microsiemens/micromhos, while the ppm/TDS

equivalents are based on our 442 Natural Water™, and NaCl

standards. All standards are within ±1.0% of reference solutions.

These same buffers may be used for calibration and recertification

of your Myron L or other handheld instrument.

a. Potassium Chloride (KCl)

The concentrations of these reference solutions are calculated

from data in the International Critical Tables, Vol. 6.

b. 442 Natural Water™

442 Natural Water Standard Solutions are based on the following

salt proportions: 40% sodium sulfate, 40% sodium bicarbonate,

and 20% sodium chloride which represent the three predominant

components “anions” in freshwater. This salt ratio has conductivity

characteristics approximating fresh natural waters and was

developed by the Myron L Company over three decades ago. It is

used around the world for measuring both conductivity and TDS in

drinking water, ground water, lakes, streams, etc.

c. Sodium Chloride (NaCl)

This is especially useful in sea water mix applications, as sodium

chloride is its major salt component. Most Myron L standard

solution labels show the ppm NaCl equivalent to the conductivity

and to ppm 442 values.

2.

pH and ORP BUFFER SOLUTIONS

For your pH and/or ORP Monitor/controller. Your pH and ORP

Monitor/controller requires the same care and calibration as your

Conductivity/TDS Monitor/controller. With the following Myron L

Company NIST traceable buffer solutions it is possible for you to

maintain the accuracy required in your process.

NOTE: pH and/or ORP Monitor/controllers ORP ZERO is

calibrated using 7pH buffer. SPAN may be calibrated using any

high quality ORP mV solution.

These same buffers may be used for calibration and recertification

of your Myron L or other handheld instrument.

a. pH Buffer Solutions

pH buffers are available in pH values of 4, 7 and 10. Myron L

Company buffer solutions are traceable to NIST certified pH

references and are color-coded for instant identification. They are

also mold inhibited and accurate to within ±0.01 pH units @ 25°C.

Order 4, 7 or 10 Buffer.

b. pH Sensor Storage Solution

Myron L Storage Solution prolongs the life of the pH sensor. It is

available in quarts and gallons. Order SSQ or SSG.

C.1.a. Potassium Chloride Standards

KCl-18

KCl-180

KCl-1800

KCl-18,000

KCl-70

KCl-700

KCl-7000

KCl-70,000

C.1.b. 442 Natural Water Standards

442-15

442-150

442-1500

442-15,000

442-30

442-300

442-3000

442-30,000

442-500

442-1000

C.1.c. Sodium Chloride Standards

NaCl-12.5

NaCl-13.4

NaCl-14.0

C.2.a. pH Buffer Solutions

pH4.0

pH7.0

pH10.0

C.2.b. pH/ORP Storage Solution

SSQ - Quart/liter

SSG - Gallon

Customer Standard Solutions available by special order @

$90.00 per Gallon. Contact us with your request.

Order example: 442-3000QT

KCl-7000GAL

24