beautypg.com

Myron L CTCIITD User Manual

Page 26

background image

3. TRANSFER STANDARD METHOD

For maximum accuracy of any Monitor/controller, the transfer

standard method should be utilized. Instead of removing the

entire Monitor/controller and sensor, and either returning it to

the manufacture or sending it to a third party laboratory for

recertification, the transfer standard allows quick recertification

and return to service — less down time. While being the most

accurate method it is also very easy to perform, and may be used

to calibrate ANY manufacture Monitor/controller — resistivity or

conductivity/TDS. This method still has the benefit of third party

verification, if so desired.

1. A high quality hand-held instrument, one capable of

accurate conductivity/TDS readings such as the Myron L

Ultrameter™, is calibrated using a standard solution, or if

so desired, sent to a third party laboratory for calibration

& certification. Preferably the standard solution should

be as close as practicable to the area of the Monitor/

controller operation, i.e. for a conductivity Monitor/

controller that is operating between 1500 and 2000µS,

the recommended Myron L standard solution is KCl-

1800.

2. Once the Ultrameter is calibrated, the process water is

measured.

3. Finally, the Monitor/controller is calibrated/adjusted to

match the value measured by the Ultrameter by adjusting

the Main CALibration control shown in figure V.A.1.

4. Turn power OFF.

5. Re-install front panel as described below in

“REASSEMBLY”.

REASSEMBLY

1. Carefully reinstall the front panel.

2. Reinstall the clear cover.

3. Reinstall the four (4) screws and tighten.

4. If so desired, install corner covers.

5. To operate, turn power ON.

C.

PREVENTIVE CARE

The Myron L Company recommends that the following Preventive

Care procedures be observed.

1. Try to prevent exposure to excessive heat and moisture.

2. The main AC power source should be protected against

excessive voltage “spikes.”

3. Take care not to damage the Monitor or sensor during

handling.

NOTE: Daily, weekly or monthly maintenance schedules are

based upon the frequency of use and the severity of the CTCII

Chemical Treatment Controllers environment and operating

conditions.

4. Repeat the Check-Out procedures to verify satisfactory

operation and/or isolate possible troubleshooting

symptoms.

5. Check all cable connections to ensure that they are free

of moisture and contamination.

6. Inspect and replace damaged component boards and

cable assemblies.

7. Periodically remove, clean and inspect the Sensor.

CLEANING - clean clear cover with soap and water, and soft non-

abrasive cloth.

22

This manual is related to the following products: